Computational hemodynamic indices to identify Transcatheter Aortic Valve Implantation degeneration / L. Crugnola, C. Vergara, L. Fusini, I. Fumagalli, G. Luraghi, A. Redaelli, G. Pontone. - In: COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE. - ISSN 0169-2607. - 259:(2025 Feb), pp. 108517.1-108517.13. [10.1016/j.cmpb.2024.108517]
Abstract:
Background and Objectives: Structural Valve Deterioration (SVD) is the main limiting factor to the long-term durability of the bioprosthetic valves used for Transcatheter Aortic Valve Implantation (TAVI), a minimally invasive technique for the treatment of severe aortic stenosis. The aim of this retrospective study is to perform patient-specific computational analyses of blood dynamics shortly after TAVI to identify hemodynamic indices that correlate with a premature onset of SVD which is detected at 5-10 years long-term follow-up exam after TAVI. Methods: The study population comprises fourteen patients: seven cases with SVD at long-term follow-up were identified and seven cases without SVD were randomly extracted from the same cohort. Starting from pre-operative CT images, we created trustworthy post-TAVI scenarios by virtually inserting the bioprosthetic valve (stent and leaflets) and we qualitatively validated such virtual scenarios against post-TAVI CT scans, when available. We then performed numerical simulations imposing personalized inlet conditions based on patient-specific Echo Doppler cardiac output measurements and the numerical results were post-processed to identify suitable hemodynamics indices with the aim of discriminating between the SVD and non-SVD groups of patients. In particular, differences in terms of each individual index were evaluated using a Wilcoxon rank-sum test. Moreover, we defined three synthetic scores, based on suitably scaled hemodynamic indices of stress and vorticity, evaluated in different contexts: on the leaflets, in the ascending aorta, and in the whole domain. Results: We found that the hemodynamic index related to leaflets’ OSI individually shows statistically significant differences (p=0.007) between the SVD and non-SVD groups. Moreover, our proposed synthetic scores are able to clearly isolate the SVD group both in a two-dimensional space given by the aorta and leaflets scores and by only considering the global synthetic score. Conclusion: The results of this computational study suggest that blood dynamics may play an important role in creating the conditions that lead to SVD. Moreover, the proposed synthetic scores could provide further indications for clinicians in assessing and predicting TAVI valves’ long-term performance.
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
Transcatheter Aortic Valve Implantation; structural valve deterioration; long-term durability computational fluid dynamics; hemodynamic indices
Elenco autori:
L. Crugnola, C. Vergara, L. Fusini, I. Fumagalli, G. Luraghi, A. Redaelli, G. Pontone
Autori di Ateneo:
FUMAGALLI BIANCA SOFIA IRENE ( autore )
PONTONE GIANLUCA ( autore )
Link alla scheda completa:
https://air.unimi.it/handle/2434/1159058
Link al Full Text:
https://air.unimi.it/retrieve/handle/2434/1159058/2820334/computational.pdf
Progetto:
PREDICTING THE OUTCOME OF ENDOVASCULAR REPAIR FOR THORACIC AORTIC ANEURYSMS: ANALYSIS OF FLUID DYNAMIC MODELING IN DIFFERENT ANATOMICAL SETTINGS AND CLINICAL VALIDATION
  • Aree Di Ricerca

Aree Di Ricerca

Settori


Settore MEDS-07/B - Malattie dell'apparato cardiovascolare
  • Informazioni
  • Assistenza
  • Accessibilità
  • Privacy
  • Utilizzo dei cookie
  • Note legali

Realizzato con VIVO | Progettato da Cineca | 25.11.5.0