Plaque quantification by coronary computed tomography angiography using intravascular ultrasound as a reference standard: a comparison between standard and last generation computed tomography scanners
Articolo
Data di Pubblicazione:
2019
Citazione:
Plaque quantification by coronary computed tomography angiography using intravascular ultrasound as a reference standard: a comparison between standard and last generation computed tomography scanners / E. Conte, S. Mushtaq, G. Pontone, L. Li Piani, P. Ravagnani, S. Galli, C. Collet, J. Sonck, L. Di Odoardo, M. Guglielmo, A. Baggiano, D. Trabattoni, A. Annoni, M.E. Mancini, A. Formenti, G. Muscogiuri, M. Magatelli, F. Nicoli, C. Poggi, C. Fiorentini, A.L. Bartorelli, M. Pepi, P. Montorsi, D. Andreini. - In: EUROPEAN HEART JOURNAL. CARDIOVASCULAR IMAGING. - ISSN 2047-2404. - (2019 May 15). [Epub ahead of print] [10.1093/ehjci/jez089]
Abstract:
Aims
The emerging role of coronary computed tomography angiography (CCTA) as a non-invasive tool for atherosclerosis evaluation is supported by data reporting a good correlation between CCTA and intravascular ultrasound
(IVUS) for plaque volume quantification. Aim of the present study was to evaluate whether a last generation CTscanner may improve coronary plaque volume assessment using IVUS as standard-of-reference.
Methods and results
From a registry of 1915 consecutive, all-comers, patients who underwent a clinically indicated IVUS evaluation we
enrolled 59 patients who underwent CCTA with a 64-slice CT (Group 1) and 59 patients who underwent CCTA
with whole-heart coverage CT scanner (Group 2). Patients who underwent CCTA with unfavourable heart rhythm
were not excluded from the analysis. Image quality (4-point Likert scale) focused on plaque analysis was evaluated.
Plaque volume quantification by CCTA was compared to IVUS. No difference in clinical characteristics was found
between Group 1 and Group 2. Plaque volume quantification by CCTA was considered not feasible in 11 plaques
of Group 1 and in 4 plaques of Group 2 (P = 0.09). Higher correlation for plaque volume quantification by CCTA
vs. IVUS was demonstrated in Group 2 when compared with Group 1 (r = 0.9888 vs. 0.9499; P < 0.0001). The
Bland–Altman analysis showed plaque volume overestimation by CCTA of 11.9 mm3 in Group 1 and 4 mm2 in
Group 2 (P < 0.001). Effective radiation dose of CCTA was significantly lower in Group 2 vs. Group 1 (2.7 ± 0.9 vs.
8.1 ± 3.6 mSv, respectively; P < 0.001).
Conclusions
CCTA using a new scanner generation showed to be an accurate non-invasive tool to assess and quantify coronary
plaque volume.
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
atherosclerosis; coronary computed tomography angiography; coronary plaque analysis; coronary plaque volume; intravascular ultrasound; whole-heart coverage CT
Elenco autori:
E. Conte, S. Mushtaq, G. Pontone, L. Li Piani, P. Ravagnani, S. Galli, C. Collet, J. Sonck, L. Di Odoardo, M. Guglielmo, A. Baggiano, D. Trabattoni, A. Annoni, M.E. Mancini, A. Formenti, G. Muscogiuri, M. Magatelli, F. Nicoli, C. Poggi, C. Fiorentini, A.L. Bartorelli, M. Pepi, P. Montorsi, D. Andreini