Pocket-size ultrasound device in cholelithiasis: diagnostic accuracy and efficacy of short-term training

Articolo
Data di Pubblicazione:
2018
Citazione:
Pocket-size ultrasound device in cholelithiasis: diagnostic accuracy and efficacy of short-term training / M. Del Medico, A. Altieri, G. Carnevale-Maffè, P. Formagnana, F. Casella, M. Barchiesi, M. Bergonzi, C. Vattiato, G. Casazza, C. Cogliati. - In: INTERNAL AND EMERGENCY MEDICINE. - ISSN 1828-0447. - 13:7(2018 Oct), pp. 1121-1126.
Abstract:
Pocket-size ultrasound devices (PSUD) are now widely available becoming a useful tool for diagnostic and therapeutic purposes. We aim to investigate the accuracy of PSUD in diagnosing cholelithiasis as compared to traditional ultrasonography. Moreover, we tested the reliability of PSUD when performed by inexperienced internal medicine residents after a short-term training. We consecutively enrolled inpatients and outpatients referred to undergo abdominal ultrasonography for signs or symptoms of gallbladder diseases in two different hospitals. Every patient underwent two independent examinations with PSUD by both expert (EXPPSUD) and nonexpert operators (N-EXPPSUD), and a conventional examination with traditional abdominal ultrasound (AUS). Every naive operator underwent a short-term training with a 2-h theoretical lesson, and a practical training focused on gallbladder under expert operator supervision. Overall, 146 patients were consecutively enrolled. Considering conventional AUS as the reference standard, sensitivity and specificity of EXPPSUDwere, respectively, 93.75 and 100%. Sensitivity and specificity of N-EXPPSUDwere, respectively, 75 and 91.25%. Nevertheless, when considering outpatients, PSUD has a high diagnostic accuracy even when performed by N-EXPPSUDwith sensitivity of 93% and specificity of 88%. PSUD is a reliable tool for the diagnosis of cholelithiasis when used by expert operators potentially reducing the need for further diagnostic tests. It can even be successfully used by non-expert operators in outpatients setting after a short focussed training.
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
Cholelithiasis; Pocket size ultrasound device; Short-term training; Ultrasound; Internal Medicine; Emergency Medicine
Elenco autori:
M. Del Medico, A. Altieri, G. Carnevale-Maffè, P. Formagnana, F. Casella, M. Barchiesi, M. Bergonzi, C. Vattiato, G. Casazza, C. Cogliati
Autori di Ateneo:
CASAZZA GIOVANNI ( autore )
COGLIATI CHIARA BEATRICE ( autore )
Link alla scheda completa:
https://air.unimi.it/handle/2434/595170
  • Aree Di Ricerca

Aree Di Ricerca

Settori (2)


Settore MED/01 - Statistica Medica

Settore MED/09 - Medicina Interna