A validation study of a new nasogastric polyfunctional catheter

Academic Article
Publication Date:
2011
Citation:
A validation study of a new nasogastric polyfunctional catheter / D. Chiumello, E. Gallazzi, A. Marino, V. Berto, C.M. Mietto, B. Cesana, L. Gattinoni. - In: INTENSIVE CARE MEDICINE. - ISSN 0342-4642. - 37:5(2011 May), pp. 791-795.
abstract:
PURPOSE: Pleural and abdominal pressure are clinically estimated by measuring the esophageal and bladder or intragastric pressure (IGP), respectively. A new nasogastric polyfunctional catheter is now commercially available, equipped with two balloons in the lower and distal part; this catheter allows simultaneous esophageal pressure (Pes) and IGP measurements and can be also used to feed the patient. We compared the Pes and IGP measured using this new device with those obtained with a standard balloon catheter taken as gold standard.
METHODS: Twenty-four intubated patients requiring ventilator support (mean age 64.3 ± 16.8 years, body mass index 25.3 ± 3.0 kg/m(2), and PaO(2)/FiO(2) 280.8 ± 123.4 mmHg) were enrolled. Esophageal pressure and IGP were measured with the new nasogastric polyfunctional catheter (Nutrivent, Sidam, Italy) and with a standard balloon catheter (Smart Cath Viasys, USA). The Smart Cath was first inserted in the stomach and then retracted to the esophagus to measure IGP and Pes, respectively. In each patient two paired measurements were averaged.
RESULTS: In the Bland-Altman analysis, the bias and agreement bands for Pes, ΔPes (computed as the difference of esophageal pressure between end-inspiration and expiration), and IGP were -0.25 (-2.65 to +2.15), 0.0 (-0.9 to +0.9), and -0.45 (-2.85 to + 1.95) cmH(2)O, respectively. No side effects or complications were recorded.
CONCLUSIONS: The new polyfunctional catheter showed a clinically acceptable validity in recording esophageal and intragastric pressure. This device should help physicians to better individualize the clinical patient management.
IRIS type:
01 - Articolo su periodico
Keywords:
Nasogastric catheter ; esophageal pressure ; intragastric pressure ; ARDS
List of contributors:
D. Chiumello, E. Gallazzi, A. Marino, V. Berto, C.M. Mietto, B. Cesana, L. Gattinoni
Authors of the University:
CHIUMELLO DAVIDE ALBERTO ( author )
Link to information sheet:
https://air.unimi.it/handle/2434/155950
  • Research Areas

Research Areas

Concepts


Settore MED/41 - Anestesiologia
  • Guide
  • Help
  • Accessibility
  • Privacy
  • Use of cookies
  • Legal notices

Powered by VIVO | Designed by Cineca | 26.7.0.0