High-performance liquid chromatographic method for measuring total plasma homocysteine levels
Articolo
Data di Pubblicazione:
1998
Citazione:
High-performance liquid chromatographic method for measuring
total plasma homocysteine levels / I. Fermo, C. Arcelloni, G. Mazzola, A. D'Angelo, R. Paroni. - In: JOURNAL OF CHROMATOGRAPHY B. BIOMEDICAL SCIENCES AND APPLICATIONS. - ISSN 1387-2273. - 719:1-2(1998 Nov 20), pp. 31-36.
Abstract:
We have modified a high-performance liquid chromatographic (HPLC) procedure based on SBD-F (ammonium-7-
fluorobenzo-2-oxa-1,3-diazole-4-sulphonate) pre-column derivatization to obtain an assay that is useful for routine clinical
total plasma homocysteine (tHcy) analysis. The introduction of easily handled sodium borohydride instead of the traditional
tri-n-butylphosphine in dimethylformamide as a reductant and a 14-min run-time using basic isocratic HPLC equipment are
the more notable advantages. The addition of mercaptopropionylglycine as an internal standard contributed to improvements
in the reproducibility of the assay, yielding within- and between-run precisions of 1.9 and 4% (C.V.), respectively. Reference
values for fasting tHcy were 7.6562.3 and 8.962.4 mmol/ l, while post-methionine load gave tHcy levels of 19.965.5 and
26.865.5 mmol/ l, for women and men, respectively (n540).
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
derivatization, LC ; homocysteine ; methionine ; amino acids
Elenco autori:
I. Fermo, C. Arcelloni, G. Mazzola, A. D'Angelo, R. Paroni
Link alla scheda completa: