Data di Pubblicazione:
2013
Citazione:
Standardizzazione della misura e traguardi analitici per l’emoglobina glicata / F. Braga, M. Panteghini. - In: BIOCHIMICA CLINICA. - ISSN 0393-0564. - 37:2(2013), pp. 100-107.
Abstract:
Measurement standardization and analytical goals for glycated hemoglobin. Glycated hemoglobin (HbA1c)
plays a key role in diagnosing diabetes and monitoring the glycemic state. To guarantee the reliability of its
measurement at global level, IFCC has defined a reference measurement system, based on the definition of the
measurand as hemoglobin molecules having a special hexapeptide in common, which is the stable adduct of glucose
to the N-terminal valine of the hemoglobin β-chain. In addition to the traceability of HbA1c results to the reference system, the establishment of analytical goals to make HbA1c measurements clinically reliable becomes crucial.
However, allowable goals will depend on the assay specificity (i.e., selectivity) and, consequently, on units in which HbA1c results are expressed [mmol/mol for IFCC-aligned systems or % for National Glycohemoglobin Standardization
Program (NGSP) converted numbers]. In this regard, analytical goals derived from biologic variability studies in which the determination of HbA1c has been carried out by an assay providing the same selectivity for the measurand as defined by the IFCC are recommended. Only these targets should be used for evaluating the performance of
commercial assays traceable to the IFCC system and of clinical laboratories using them through appropriately
structured quality controls. Analytical systems following different calibration hierarchies (e.g., the NGSP-aligned
assays) will require different analytical goals.
Tipologia IRIS:
01 - Articolo su periodico
Elenco autori:
F. Braga, M. Panteghini