Surface-activated chemical ionization ion trap mass spectrometry in the analysis of amphetamines in diluted urine samples
Articolo
Data di Pubblicazione:
2004
Citazione:
Surface-activated chemical ionization ion trap mass spectrometry in the analysis
of amphetamines in diluted urine samples / S. Cristoni, L. Rossi Bernardi, P.Gerthoux, E. Gonella , P. Mocarelli. - In: RAPID COMMUNICATIONS IN MASS SPECTROMETRY. - ISSN 0951-4198. - 18:16(2004), pp. 1847-1855.
Abstract:
A new ionization method, named surface-activated chemical ionization (SACI), was
employed for the analysis of five amphetamines (3,4-methylenedioxyamphetamine
(MDA), 3,4-methylenedioxymethamphetamine (MDMA), 3,4-methylenedioxyethylamphetamine (MDE), amphetamine and methamphetamine) by
ion trap mass spectrometry. The results so obtained have been compared with those achieved by using atmospheric pressure chemical ionization (APCI) and electrospray ionization (ESI) using the same instrument, clearly showing that SACI is the most sensitive of the three. The limit of detection and linearity range for SACI were compared with those obtained using APCI and ESI, showing that the new SACI approach provides the best results for both criteria. SACI was used to analyze MDA, MDMA MDE, amphetamine and methamphetamine in four urine
samples, and the quantitation results are compared with those achieved using
ESI.
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
Amphetamines
Central Nervous System Stimulants Ions
Elenco autori:
S. Cristoni, L. Rossi Bernardi, P.Gerthoux, E. Gonella , P. Mocarelli