Data di Pubblicazione:
1983
Citazione:
A new method to measure the hemoglobin oxygen saturation by the oxygen electrode / M. Samaja, E. Rovida. - In: JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS. - ISSN 0165-022X. - 7:2(1983 Feb), pp. 143-152. [10.1016/0165-022X(83)90047-7]
Abstract:
We describe a new polarographic method to measure the haemoglobin oxygen saturation in whole blood, employing up to 10 μl of sample in a standard case. The measurement is done in an anaerobic staineless-steel cuvette (1 ml) recording three oxygen tension values: (1) that of an air-equilibrated buffer before the addition of the sample; (ii) that after the addition of the sample; and (iii) that after the addition of an oxidant. The haemoglobin oxygen saturation is then calculated from the three oxygen tension values, the volume of the reagents, and solubility coefficient of oxygen. This method is simple, inexpensive and accurate, and correlates well with other standard methods.
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
haemoglobin oxygen saturation; oxygen tension; polrographic methods
Elenco autori:
M. Samaja, E. Rovida
Link alla scheda completa: