Evaluation of commutability of the ERM - DA 470k reference material for two albumin assays using immunochemical principles
Poster
Data di Pubblicazione:
2011
Citazione:
Evaluation of commutability of the ERM - DA 470k reference material for two albumin assays using immunochemical principles / I. Infusino, F. Braga, C. Valente, M. Panteghini. - In: CLINICAL CHEMISTRY AND LABORATORY MEDICINE. - ISSN 1434-6621. - 49:Suppl.1(2011 May), pp. S805-S805. ((Intervento presentato al convegno IFCC - WordLab - EuroMedLab tenutosi a Berlin nel 2011 [10.1515/CCLM.2011.530].
Abstract:
Background. To standardize the measurement of human serum proteins, a new reference material (RM)(IRMM ERM-DA 470k/
IFCC) has recently been released. We evaluatedthe commutability of this RM for two immunochemical albumin methods based
on turbidimetry (Roche Cobas Tina-quant Gen. 2) and nephelometry (Delta Radim) principles.
Methods. We measured albumin concentrations on the two systems in20 human serum samples (albumin concentrations, 16.3
to 42.3 g/L) and in the RM in duplicate in two different runs on the same day. Biological samples were collected, aliquoted and
stored(£24 h) at +4°C until their use. Manufacturer’s control materials were used to validate analytical runs. The RM commutability
was estimated from data of regression analysis using the 95% prediction interval (95PI) and multiples of the standard error of
regression (Sy-x), according to the CLSI C53-A document.
Results. The RM results were contained within the 95PIbasedon the results for the native clinical samples. In addition, using an
acceptance criterion for commutability of ±2 times the experimental Sy-x (±3.135), the relative residual for the RM (-2.107) was
within the acceptable range. Albumin values for RM on both analytical systems (34.6 g/L for turbidimetry and 34.2 g/L for nephelometry)
were, however, markedly lower than expected (certified value±uncertainty, 37.2±1.2 g/L).
Conclusions. The new RM is commutable between the evaluated immunochemical assays and can be used as a basis to maintain
their traceability to higher-order references. However, some inconsistency in the assignment of values to the manufacturer’s
working calibrators is likely and should be verified.
IFCC) has recently been released. We evaluatedthe commutability of this RM for two immunochemical albumin methods based
on turbidimetry (Roche Cobas Tina-quant Gen. 2) and nephelometry (Delta Radim) principles.
Methods. We measured albumin concentrations on the two systems in20 human serum samples (albumin concentrations, 16.3
to 42.3 g/L) and in the RM in duplicate in two different runs on the same day. Biological samples were collected, aliquoted and
stored(£24 h) at +4°C until their use. Manufacturer’s control materials were used to validate analytical runs. The RM commutability
was estimated from data of regression analysis using the 95% prediction interval (95PI) and multiples of the standard error of
regression (Sy-x), according to the CLSI C53-A document.
Results. The RM results were contained within the 95PIbasedon the results for the native clinical samples. In addition, using an
acceptance criterion for commutability of ±2 times the experimental Sy-x (±3.135), the relative residual for the RM (-2.107) was
within the acceptable range. Albumin values for RM on both analytical systems (34.6 g/L for turbidimetry and 34.2 g/L for nephelometry)
were, however, markedly lower than expected (certified value±uncertainty, 37.2±1.2 g/L).
Conclusions. The new RM is commutable between the evaluated immunochemical assays and can be used as a basis to maintain
their traceability to higher-order references. However, some inconsistency in the assignment of values to the manufacturer’s
working calibrators is likely and should be verified.
Tipologia IRIS:
01 - Articolo su periodico
Elenco autori:
I. Infusino, F. Braga, C. Valente, M. Panteghini
Link alla scheda completa: