Data di Pubblicazione:
2010
Citazione:
Start to end simulation of a Thomson source for mammography / P. Oliva, U. Bottigli , L.A. Gizzi, A. Bacci, S. Betti, C.A. Cecchetti, M. Ferrario, A. Gamucci, A. Giulietti, D. Giulietti, L. Labate, T. Levato, V. Petrillo, L. Serafini, P. Tomassini, C. Vaccarezza. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 615:1(2010), pp. 93-99.
Abstract:
Thomson scattering X-ray sources have many features which are of relevance for several applications: the main one is the capability to produce intense, quasi-monochromatic, tunable X-ray beams, after collimation, still with a reasonably small size apparatus. Applications to medical physics are straightforward, in particular in mammography where dose control in screening programs is the main relevant issue. An optimal choice of the X-ray energy to image the breast will result in a best image quality and hence will lead to a dose reduction.
A Thomson scattering source is presently under development at the Frascati National Laboratories (LNF) of INFN (Istituto Nazionale di Fisica Nucleare).
A complete simulation of the source including electron beam, laser beam, Thomson interaction and X-ray imaging is presented. The X-rays are generated in the energy range suitable for mammography and used to generate images of a mammographic phantom.
Image quality is evaluated in terms of dose efficiency and compared to those obtained by monochromatic beams and conventional X-ray tubes.
A Thomson scattering source is presently under development at the Frascati National Laboratories (LNF) of INFN (Istituto Nazionale di Fisica Nucleare).
A complete simulation of the source including electron beam, laser beam, Thomson interaction and X-ray imaging is presented. The X-rays are generated in the energy range suitable for mammography and used to generate images of a mammographic phantom.
Image quality is evaluated in terms of dose efficiency and compared to those obtained by monochromatic beams and conventional X-ray tubes.
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
Mammography; Thomson scattering
Elenco autori:
P. Oliva, U. Bottigli , L.A. Gizzi, A. Bacci, S. Betti, C.A. Cecchetti, M. Ferrario, A. Gamucci, A. Giulietti, D. Giulietti, L. Labate, T. Levato, V. Petrillo, L. Serafini, P. Tomassini, C. Vaccarezza
Link alla scheda completa: