Cyclic intermittent aortic regurgitation of a mechanical bileaflet aortic valve prosthesis : diagnosis and clinical implications
Articolo
Data di Pubblicazione:
2007
Citazione:
Cyclic intermittent aortic regurgitation of a mechanical bileaflet aortic valve prosthesis : diagnosis and clinical implications / C.A. Galli, M. Muratori, P. Montorsi, F. Barili, G. Polvani, M. Pepi. - In: JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY. - ISSN 0894-7317. - 20:11(2007 Nov), pp. 1315.e5-1315.e8.
Abstract:
We describe a rare case of intermittent dysfunction of a bileaflet mechanical aortic valve prosthesis. Subvalvular pannus overgrowth on the inflow aspect of the prosthesis impeded normal closure of one leaflet causing severe aortic regurgitation, which occurred intermittently. The complementary role of transthoracic and transesophageal echocardiography and cine-fluoroscopy in the diagnosis of this case is described.