Multi-institutional application of Failure Mode and Effects Analysis (FMEA) to CyberKnife Stereotactic Body Radiation Therapy (SBRT)
Articolo
Data di Pubblicazione:
2015
Citazione:
Multi-institutional application of Failure Mode and Effects Analysis (FMEA) to CyberKnife Stereotactic Body Radiation Therapy (SBRT) / I. Veronese, E. De Martin, A.S. Martinotti, M.L. Fumagalli, C. Vite, I. Redaelli, T. Malatesta, P. Mancosu, G. Beltramo, L. Fariselli, M.C. Cantone. - In: RADIATION ONCOLOGY. - ISSN 1748-717X. - 10:1(2015 Jun 13), pp. 132.1-132.10. [10.1186/s13014-015-0438-0]
Abstract:
Background A multidisciplinary and multi-institutional working group applied the Failure Mode and Effects Analysis (FMEA) approach to assess the risks for patients undergoing Stereotactic Body Radiation Therapy (SBRT) treatments for lesions located in spine and liver in two CyberKnife® Centres. Methods The various sub-processes characterizing the SBRT treatment were identified to generate the process trees of both the treatment planning and delivery phases. This analysis drove to the identification and subsequent scoring of the potential failure modes, together with their causes and effects, using the risk probability number (RPN) scoring system. Novel solutions aimed to increase patient safety were accordingly considered. Results The process-tree characterising the SBRT treatment planning stage was composed with a total of 48 sub-processes. Similarly, 42 sub-processes were identified in the stage of delivery to liver tumours and 30 in the stage of delivery to spine lesions. All the sub-processes were judged to be potentially prone to one or more failure modes. Nineteen failures (i.e. 5 in treatment planning stage, 5 in the delivery to liver lesions and 9 in the delivery to spine lesions) were considered of high concern in view of the high RPN and/or severity index value. Conclusions The analysis of the potential failures, their causes and effects allowed to improve the safety strategies already adopted in the clinical practice with additional measures for optimizing quality management workflow and increasing patient safety.
Tipologia IRIS:
01 - Articolo su periodico
Keywords:
FMEA; SBRT; CyberKnife; Tracking; Liver; Spine
Elenco autori:
I. Veronese, E. De Martin, A.S. Martinotti, M.L. Fumagalli, C. Vite, I. Redaelli, T. Malatesta, P. Mancosu, G. Beltramo, L. Fariselli, M.C. Cantone
Link alla scheda completa: